Description
DMP 10-40
- Universal instrument series for tactile coating thickness measurement thanks to an incredibly wide probe portfolio
- Measured value memory: 250,000 measured values in 2,500 batches (DMP30/40) or 10,000 measured values in 1 batch (DMP10/20)
- Robust aluminum housing with IP64 protection, Gorilla Glass and soft bumpers
- Exchangeable Li-Ion battery for > 24 h operating time
- Easy data transfer via USB-C and Bluetooth
- Live feedback via light, sound and vibration for limit monitoring
- Cal Check function to verify your currently used calibration
- Digital and analog probes available for a wide range of applications, with the F-adapter you still have access to the entire Fischer probe portfolio
- Available with the new Tactile Suite software
ISOSCOPE®
Measurement of electrically non-conductive layers on non-magnetizable, electrically conductive base materials.
Application examples:
- NC/NF: Paint, varnish or plastic on aluminum, copper or brass
- NC/NF: Anodized coatings on aluminum
Measurement method:
- Amplitude sensitive eddy current method
DELTASCOPE®
Measurement of non-magnetizable layers on magnetizable base materials.
Application examples:
- NC/Fe: Paint, varnish or plastic on steel or iron
- NF/Fe: Zinc, chrome or copper on steel or iron
Measurement method:
- Magnetic inductive measuring method
DUALSCOPE®
Measurement of electrically non-conductive layers on non-magnetizable, electrically conductive base materials and measurement of non-magnetizable layers on magnetizable base materials.
Application examples:
- NC/NF: Paint, varnish, plastic on aluminum, copper, brass
- NC/NF: Anodized coatings on aluminum
- NC/Fe: Paint, varnish or plastic on steel or iron
- NF/Fe: Zinc, chrome or copper on steel or iron
Measurement method:
- Amplitude-sensitive eddy current method and magnetic inductive measurement method