MAHR Precimar ULM 520 S-E Calibration measuring instruments
Technical Data
Technical data
ULM 520 S-E
|
ULM 1000 S-E
|
|
Device dimensions
|
1080 x 380 x 480
|
1500 x 380 x 480
|
Measuring range for outer measurement [mm]
|
0 to 520
|
0 to 1025
|
Measuring range for inner measurement [mm]
|
0.5 to 365
|
0.5 to 870
|
Direct measuring range [mm]
|
Outer measurement: 0 to 520
Inner measurement: 0.5 to 365 |
Outer measurement: 0 to 1025
Inner measurement: 0.5 to 870 |
Measuring uncertainty MPEE1 (L in mm) [µm]
|
only with ABBE measuring element: MPE E1 ? (0.09+L/2000)
with base bed measuring system: MPE E1 ? (0.6+L/1000) |
only with ABBE measuring element: MPE E1 ? (0.09+L/2000)
with base bed measuring system: MPE E1 ? (0.6+L/1000) |
Measuring forces [N]
|
0.2 ; 1.0 to 4.5 ; 11
|
0.2 ; 1.0 to 4.5 ; 11
|
Repeatability [µm]
|
With Abbe measuring element: ≤ 0.05
With base bed measuring system: ≤ 0.2 |
With Abbe measuring element: ≤ 0.05
With base bed measuring system: ≤ 0.2 |
Device length [mm]
|
1080
|
1500
|
Mass [kg]
|
160
|
215
|
Applications
Calibration of:
- Plain plug gages and gage rings
- Setting gage rings
- Snap gages
- Spherical gage blocks, internal micrometers
- Gage blocks
- Thread gages
- Taper thread gages
- Gear gages
- Taper gages
- Dial indicators
- Dial comparators
- 2-point bore gages
- Outside micrometers
- 2-point inside micrometers
Accessories
- Large number of accessory kits and modular components for completing a wide variety of measuring tasks, including the measurement of:
- Thread gages
- Taper gages
- Conical thread gages
- Gears
- Factory calibration or DAkkS/DKD calibration can be additionally offered for this measuring station